Mass range: at least 6amu, the maximum mass is not less than 2980 amu
Resolution(full mass range): unit mass resolution (full scan)
MRM sensitivity:
electrospray ionization ESI: positive ion 1 pg, Reserpine direct injection, MRM analysis609/195, signal to noise ratio > 2500:1 (RMS)
Negative ion 1pg chloramphenicol direct injection, MRM analysis 312/152, signal to noiseratio S/N>1400:1
APCI:1pg Reserpine direct injection, MRM analysis measurement 609/195, signal tonoise ratio >500:1 (peak to peak ratio)
Detection limit of mass spectrometry: Clenbuter detection limit ≤0.01ppb; chloramphenicol detection limit ≤0.01ppb
Scanning speed: maximum >2300 amu/sec
Dynamic linear range of data acquisition: >4*106
Linear Accelerated High Pressure Collision Chamber Design: Pressure greater than 7mtor, improves ion transmission efficiency, improves sensitivity, eliminates memory effects and cross-contamination
Quantitative analysis reproducibility:
low concentration: chloramphenicol 0.1 ppb, injection volume 20 μl, N=5, CV <3%;
High concentration: chloramphenicol 2ppb, injection volume 20μl, N=5, CV<3%;
5ppb;50ppb glycerol and cholesterol were injected 5 times respectively RSD<1%
One injection, regardless of time window: 300 compound MRM quantification andlibrary search simultaneously complete no loss of sensitivity, with actualspectral support.
1. API4000 mass spectrometer host (ESI and APCI ion source)
2. Pre-stage vacuum pump for mass spectrometry
3. Supporting software and computer (Analyst 1.6.3 system software)
4. Supporting nitrogen generator (optional)
5. UPS uninterrupted protection power supply (optional)